These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of N-desmethyldiazepam in plasma by gas chromatography with an internal standard.
    Author: Viala A, Cano JP, Reynier JP, Rispe R.
    Journal: J Chromatogr; 1978 Jan 11; 147():349-57. PubMed ID: 340466.
    Abstract:
    N-Desmethyldiazepam in plasma has been determined by gas chromatography with electron capture detection, using N-desmethyltetrazepam as internal standard and a column filled Poly-A 103 on Gas-Chrom Q. No prior transformation of benzodiazepines is necessary. The sensitivity is ca.1 ng/ml and the recovery is greater than or equal to 70%. The method is particularly suitable for the pharmacokinetic study of dipotassium clorazepate: the amount of unchanged clorazepate may be estimated from the difference between the amount of N-desmethyldiazepam present initially and the total amount of N-desmethyldiazepam obtained after acidic treatment of the sample; the method is also useful for therapeutic control and for analytical toxicology.
    [Abstract] [Full Text] [Related] [New Search]