  • Title: Simultaneous liquid-chromatographic determination of prednisone and prednisolone in plasma.
    Author: Cheng MH, Huang WY, Lipsey AI.
    Journal: Clin Chem; 1988 Sep; 34(9):1897-9. PubMed ID: 3416442.
    Abstract:
    This high-performance liquid-chromatographic (HPLC) method for simultaneous determination of prednisone and its metabolite, prednisolone, in plasma is a modification of the method of Frey et al. (Clin Chem 1979;25:1944-7). Heparinized plasma (1.0 mL) with 0.1 mL of internal standard solution (11-deoxy-17-hydroxycorticosterone, 2 mg/L) is extracted with 7.0 mL of dichloromethane, then washed sequentially with 0.1 mol/L HCl, 0.1 mol/L NaOH, and deionized water, 2.0 mL each. The extract is evaporated and the residue reconstituted with 75 microL of mobile phase, methanol/H2O (40/60 by vol). Thirty microliters of this is injected onto a reversed-phase C6 column, which is eluted at 1.4 mL/min. Analytical recoveries of prednisone and prednisolone were 94-98% and 102-106%, respectively. Day-to-day precision (CV) was 3.8% for prednisone, 6.1% for prednisolone. We encountered no interference from the 21 other steroids and 25 drugs tested. This method is simple, accurate, and precise.