These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Simultaneous determination of amino acid and monoamine neurotransmitters in serum by high performance liquid chromatography coupled with precolumn derivatization]. Author: Bai J, Wang D, Liu Z, Zhang J, Liu L, Han Y. Journal: Se Pu; 2020 Aug 08; 38(8):923-928. PubMed ID: 34213184. Abstract: Using o-phthalaldehyde (OPA) as the derivatization reagent, a precolumn derivatized -high performance liquid chromatography (HPLC) method was developed for the simultaneous determination of amino acid neurotransmitters taurine (Tau), glutamic acid (Glu), glycine(Gly), and γ -aminobutyric acid (γ-GABA), as well as the monoamine neurotransmitter dopamine (DA), in serum samples. The samples and ethanol were mixed at a volume ratio of 1:2 (v/v) for protein precipitation. After centrifugation, the supernatant was withdrawn and blown to dryness using nitrogen. The residue was pre-column derivatized with OPA, and the derivatized product was isolated by gradient elution ona Luna 5u C18 column (250 mm×4.6 mm, 5 μm). Under the optimal experimental conditions, the five neurotransmitters showed good linearities (r2 ≥ 0.9866). The limits of detection were between 0.10 and 0.40 μmol/L. The spiked recoveries at different spiked levels were 87.57%-115.31%, and the RSDs were below 7.80%. This method is simple, sensitive, and it can be promised for the simultaneous detection of amino acid and monoamine neurotransmitters.[Abstract] [Full Text] [Related] [New Search]