These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of Elemental Concentrations in Lichens Using ICP-AES/MS. Author: Zhao LC, Wang L, Jiang YJ, Hu YQ, Xu CY, Wang L, Li X, Wei L, Guo XP, Liu AQ, Liu HJ. Journal: Bio Protoc; 2017 Mar 05; 7(5):e2165. PubMed ID: 34458478. Abstract: Lichens are good biomonitors for air pollution because of their high enrichment capability of atmospheric chemical elements. To monitor atmospheric element deposition using lichens, it is important to obtain information on the multi-element concentrations in lichen thalli. Because of serious air pollution, elemental concentrations in thalli of lichens from North China (especially Inner Mongolia, Hebei, Shanxi and Henan province) are often higher than those from other regions, therefore highlighting the necessity to optimize ICP-AES/MS (Inductively coupled plasma-atomic emission spectroscopy/mass spectrometry) for analyzing lichen element content. Based on the high elemental concentrations in the lichen samples, and the differences in the sensitivity and detection limits between ICP-MS and ICP-AES, we propose a protocol for analyzing 31 elements in lichens using ICP-AES/MS. Twenty-two elements (Cd, Ce, Co, Cr, Cs, Cu, K, La, Mo, Na, Ni, Pb, Rb, Sb, Sc, Sm, Sr, Tb, Th, Tl, V and Zn) can be identified by using microwave digestion- ICP-MS, and 9 elements (Al, Ba, Ca, Fe, Mg, Mn, P, S and Ti) by using ashing-alkali fusion digestion- ICP-AES.[Abstract] [Full Text] [Related] [New Search]