These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Biomechanics of a laterally placed sacroiliac joint fusion device supplemental to S2 alar-iliac fixation in a long-segment adult spinal deformity construct: a cadaveric study of stability and strain distribution.
    Author: de Andrada Pereira B, Wangsawatwong P, Lehrman JN, Sawa AGU, Lindsey DP, Yerby SA, Godzik J, Waguespack AM, Uribe JS, Kelly BP.
    Journal: J Neurosurg Spine; 2022 Jan 01; 36(1):42-52. PubMed ID: 34534964.
    Abstract:
    OBJECTIVE: S2 alar-iliac (S2AI) screw fixation effectively enhances stability in long-segment constructs. Although S2AI fixation provides a single transarticular sacroiliac joint fixation (SIJF) point, additional fixation points may provide greater stability and attenuate screw and rod strain. The objectives of this study were to evaluate changes in stability and pedicle screw and rod strain with extended distal S2AI fixation and with supplemental bilateral integration of two sacroiliac joint fusion devices implanted using a traditional minimally invasive surgical approach. METHODS: Eight L1-pelvis human cadaveric specimens underwent pure moment (7.5 Nm) and compression (400 N) tests under 4 conditions: 1) intact (pure moment loading only); 2) L2-S1 pedicle screw and rod with L5-S1 interbody fusion; 3) added S2AI screws; and 4) added bilateral laterally placed SIJF. Range of motion (ROM), rod strain, and screw-bending moment (S1 and S2AI) were analyzed. RESULTS: Compared with S1 fixation, S2AI fixation significantly reduced L5-S1 ROM in right lateral bending by 50% (0.11°, p = 0.049) and in compression by 39% (0.22°, p = 0.003). Compared with fixation ending at S1, extending fixation with S2AI significantly decreased sacroiliac joint ROM by 52% (0.28°, p = 0.02) in flexion, by 65% (0.48°, p = 0.04) in extension, by 59% (0.76°, p = 0.02) in combined flexion-extension, and by 36% (0.09°, p = 0.02) in left axial rotation. The addition of S2AI screws reduced S1 screw-bending moment during flexion (0.106 Nm [43%], p = 0.046). With S2AI fixation, posterior L5-S1 primary rod strain increased by 124% (159 μE, p = 0.002) in flexion, by 149% (285 μE, p = 0.02) in left axial rotation, and by 99% (254 μE, p = 0.04) in right axial rotation. Compared with S2AI fixation, the addition of SIJF reduced L5-S1 strain during right axial rotation by 6% (28 μE, p = 0.04) and increased L5-S1 strain in extension by 6% (28 μE, p = 0.02). CONCLUSIONS: Long-segment constructs ending with S2AI screws created a more stable construct than those ending with S1 screws, reducing lumbosacral and sacroiliac joint motion and S1 screw-bending moment in flexion. These benefits, however, were paired with increased rod strain at the lumbosacral junction. The addition of SIJF to constructs ending at S2AI did not significantly change SI joint ROM or S1 screw bending and reduced S2AI screw bending in compression. SIJF further decreased L5-S1 rod strain in axial rotation and increased it in extension.
    [Abstract] [Full Text] [Related] [New Search]