These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Exposure reduction of 96% in intraoral radiography. Author: Kircos LT, Vandre RH, Lorton L. Journal: J Am Dent Assoc; 1986 Nov; 113(5):746-50. PubMed ID: 3465788. Abstract: This study compared the diagnostic quality of E-speed film, xeroradiographic cassettes, and a screen-film imaging system used in medical radiology. The surfaces of defects simulating caries were scored for the presence and depth of the defects under ideal viewing conditions by independent observers. Receiver Operator Characteristics analysis of the data showed no significant differences regarding diagnostic potential among the three systems. The true-positive fraction was 1.00 for all systems and all operators, and the false-positive fraction was .011, .011, and .007, respectively. An exposure of 5.5 mR was used to produce images with the screen-film combination, and 150 mR was used for standard dental systems. This study showed that an exposure reduction of 96% is possible while maintaining high-quality radiographs for the diagnosis of caries.[Abstract] [Full Text] [Related] [New Search]