  • Title: Glucose oxidase@zinc-doped zeolitic imidazolate framework-67 as an effective cascade catalyst for one-step chemiluminescence sensing of glucose.
    Author: Xu X, Ma Y, Ding S, Li Y.
    Journal: Mikrochim Acta; 2021 Nov 23; 188(12):427. PubMed ID: 34812945.
    Abstract:
    A chemiluminescence (CL) sensor was constructed for the one-step determination of glucose. Glucose oxidase (GOx) was successfully encapsulated into Zn-doped zeolitic imidazolate framework-67 (Zn-ZIF-67) via a simple one-pot strategy. The as-prepared GOx@Zn-ZIF-67 nanocomposite can trigger cascade reactions of glucose oxidation to generate H2O2 and H2O2-mediated luminol reaction to give an intense CL emission. The sensor responds linearly to glucose in the 20.0-400.0 μmol·L-1 range with a limit of detection (LOD) of 4.7 μmol·L-1. Eleven replicated measurements of 200.0 μmol·L-1 glucose solution gives a relative standard deviation (RSD) of 1.7%. The sensor exhibits good selectivity and stability and was successfully applied to the determination of glucose in real human serum samples. Schematic representation of one-step determination of serum glucose with GOx@Zn-ZIF-67 nanocomposite triggering cascade reactions between luminol and glucose.