These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A Distance-Based Microfluidic Paper-Based Biosensor for Glucose Measurements in Tear Range. Author: Allameh S, Rabbani M. Journal: Appl Biochem Biotechnol; 2022 May; 194(5):2077-2092. PubMed ID: 35029790. Abstract: The prevalence of diabetes has increased over the past years. Therefore, developing minimally invasive, user-friendly, and cost-effective glucose biosensors is necessary especially in low-income and developing countries. Cellulose paper-based analytical devices have attracted the attention of many researchers due to affordability, not requiring trained personnel, and complex equipment. This paper describes a microfluidic paper-based analytical device (μPAD) for detecting glucose concentration in tear range with the naked eye. The paper-based biosensor fabricated by laser CO2; and glucose oxidase/horseradish peroxidase (GOx/HRP) enzyme solution coupled with tetramethylbenzidine (TMB) were utilized as reagents. A sample volume of 10 μl was needed for the biosensor operation and the results were observable within 5 min. The color intensity-based and distance-based results were analyzed by ImageJ and Tracker to evaluate the device performance. Distance-based results showed a linear behavior in 0.1-1.2 mM with an R2 = 0.9962 and limit of detection (LOD) of 0.1 mM. The results could be perceived by the naked eye without needing additional equipment or trained personnel in a relatively short time (3-5 min).[Abstract] [Full Text] [Related] [New Search]