These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Automated and manual quantitative assays of choriogonadotropin in serum compared.
    Author: Beinlich CJ, Carpenter RH.
    Journal: Clin Chem; 1987 Jan; 33(1):167-9. PubMed ID: 3542289.
    Abstract:
    We found the analytical performance of a rapid, automated assay of human choriogonadotropin (hCG) in serum, the Stratus hCG Fluorometric Enzyme Immunoassay, superior to a widely used manual assay for hCG (Hybritech Tandem-E hCG). The two assays were comparable in sensitivity; recovery; cross reactivity with lutropin, follitropin, and thyrotropin; and freedom from interference from hemoglobin and bilirubin. Patient-correlation studies indicated good quantitative agreement [Stratus hCG = (1.08 X Tandem hCG) - 4.3 int. units/L]. However, intra- and interassay precision was substantially better with the Stratus hCG assay, and this may allow earlier confirmation of pregnancy.
    [Abstract] [Full Text] [Related] [New Search]