These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Could S-Monovette Serum Gel Tubes be Used for Clinical Chemistry Analysis Instead of Serum Separator Tube II Advance? Author: Cat A, Ucar KT, Nurlu N. Journal: Clin Lab; 2022 Apr 01; 68(4):. PubMed ID: 35443592. Abstract: BACKGROUND: We aimed to compare the Sarstedt S-Monovette serum gel tube and the BD (Becton, Dickinson and Company) Serum Separator Tube II (SST II) Advance based on technical specifications and tests results. METHODS: One hundred and twenty volunteers were included in the technical evaluation and 42 of 120 volunteers in the clinical evaluation. Blood was collected into S-Monovette, and SST II. Twelve quality indicators (QI) were determined for technical evaluation. For clinical evaluation, 29 clinical chemistry analytes were analysed simultaneously on a Roche Cobas 6000 c501 (Roche Diagnostics, Mannheim, Germany). Calculations were made using the formula suggested by the EFLM according to the QIs. If the difference between S-Monovette and SST II was < 1%, S-Monovette was considered sufficient for relevant QI. For clinical evaluation, Passing Bablok regression analysis and Bland-Altman plots were used. Desirable bias values for comparison with mean percentage difference (MPD) were obtained from biological variation databases. RESULTS: S-Monovette tubes were found to be suitable for all QIs (difference < 1%). No significant differences were observed in analytes except lactate dehydrogenase (LDH). LDH results (U/L) obtained from the SST II were statistically significantly higher (SST II: 201 ± 42, S-Monovette: 195 ± 35, regression equation was y = 31.4 + 0.8x). The MPD of LDH (2.4%) remained within the desirable bias (3.4%); however, the 95% CI of the MPD of LDH (0.5% - 4.4%) exceeded the desirable bias. CONCLUSIONS: S-Monovette has been deemed appropriate for use in clinical chemistry analysis, as the MPD of LDH and other analytes remained within the bias limits. The LDH was considered sensitive to microhemolysis as a possible reason for the difference in LDH results.[Abstract] [Full Text] [Related] [New Search]