Title: Blood glucose monitoring by children at home: a comparison of methods. Author: Anderson DG, Gleeson M, Boulton TJ. Journal: Aust Paediatr J; 1986 Nov; 22(4):309-12. PubMed ID: 3566680. Abstract: The accuracy of measurements of blood glucose performed at home by the index children and/or parents was compared using the Boehringer Mannheim reagent strips, Ames Visidex-2 strips, an Ames Glucometer and a Boehringer Mannheim Reflolux reflectance meter. Capillary plasma samples, collected simultaneously, were later analysed by a Beckman glucose oxidase analyser. The coefficient of correlation between the self-monitored results (SMR) and the laboratory-analysed results (LAR) ranged from 0.59 to 0.92, the composite being 0.74. The mean difference between the SMR and the LAR was 1.0 mmol/l. The occurrence of potential errors of management due to incorrect SMR results was determined. On four occasions, the SMR was greater than 6 mmol/l when the LAR was less than 3 mmol/l, and on 36 occasions a discrepancy greater than 50% occurred between the two values, without resulting hypoglycaemia.[Abstract] [Full Text] [Related] [New Search]