These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [A new biopsy needle for transcutaneous histology procurement]. Author: Köhler D, Rühle KH, Costabel U, Freudenberg N, Matthys H. Journal: Schweiz Med Wochenschr; 1987 Apr 04; 117(14):527-30. PubMed ID: 3576162. Abstract: The new Vacu-Cut biopsy needle differs from previous types by its simpler manipulation and the production of a stronger vacuum to extract the tissue core. This vacuum is caused by withdrawal of a stylet through an air-tight sealing membrane. The instrument is a hollow needle with an inner cutting-edge and a diameter of 0.95 mm. 49 patients with suspected lesions of the pulmonary parenchyma were biopsied transcutaneously under x-ray monitoring. The biopsy material allowed a clearcut diagnosis in 44 patients, including 2 repeat biopsies; the success rate was thus 86% of all biopsies performed. 7 attempts (14%) did not yield useful material as the lesions were too small (diameter below 2.5 cm) or were not penetrated. In 35% of all biopsies small pneumothoraces were observed after a few hours. In two cases transitory hemoptysis occurred.[Abstract] [Full Text] [Related] [New Search]