These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Negligible Specimen Hemolysis Is Observed With Sarstedt S-Monovette Coagulation Tubes in Aspiration Mode.
    Author: Çat A, Uçar KT.
    Journal: Am J Clin Pathol; 2022 Oct 06; 158(4):494-498. PubMed ID: 35849104.
    Abstract:
    OBJECTIVES: We aimed to compare the levels of hemolysis in the blood collected using the vacuum and aspiration modes via Sarstedt S-Monovette coagulation tubes. METHODS: Forty volunteers were included in the study. Blood samples were collected using two different modes in the S-Monovette citrate tube (Sarstedt AG). Prothrombin time, active partial thromboplastin time, fibrinogen, and D-dimer analyses were performed using the STA-Compact-Max 3 analyzer (Stago). The hemolysis levels of the samples were measured by both Stago's semiquantitative hemolysis index (H-index) module and the quantitative H-index measurement of the Roche cobas 6000 (Roche Diagnostics) analyzer. RESULTS: Roche's quantitative H-index values were statistically significantly lower in the aspiration mode. No clinically significant difference was observed between coagulation test results. CONCLUSIONS: Using the S-Monovette citrate tubes can reduce spurious hemolysis and improve patient safety.
    [Abstract] [Full Text] [Related] [New Search]