These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: DOSE MEASUREMENT PRECISION OF AN RPLD-BASED EYE LENS DOSEMETER APPLICABLE TO THE MEDICAL SECTOR. Author: Kowatari M, Nagamoto K, Nakagami K, Yoshitomi H, Moritake T, Kunugita N. Journal: Radiat Prot Dosimetry; 2022 Oct 01; 198(17):1303-1312. PubMed ID: 35909091. Abstract: We demonstrate a practical calibration method and its applicability for a commercially available radiophotoluminescence dosemeter (RPLD), i.e. the GD-352M (AGC Techno Glass, Shizuoka, Japan) to eye lens dose monitoring, by performing the calibration according to the ISO recommendations. The calibration was then verified through a series of experiments. For verification of the derived calibration factor (1.21 ± 0.04, k = 1) of the RPLD, we performed standard irradiations in the ISO narrow series X-ray reference fields and the simulation measurements in the actual radiation fields in a hospital. The TLD-based commercially available dosemeters, DOSIRIS™ was also put on the ISO cylinder phantom and the RANDO phantom together with the GD-352M in the verification experiments. The personal dose equivalents Hp(3) obtained from the GD-352M and those obtained from the DOSIRIS™ were in good agreement with each other. Our results demonstrate the proper calibration of a commercially available RPLD that is applicable to the additional monitoring of the lens of the eyes for medical staff.[Abstract] [Full Text] [Related] [New Search]