These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Analytical performance and potential clinical utility of the GenMark Dx ePlex® blood culture identification gram-positive panel. Author: McCarty TP, White CM, Meeder J, Moates D, Pierce HM, Edwards WS, Hutchinson J, Lee RA, Leal SM. Journal: Diagn Microbiol Infect Dis; 2022 Nov; 104(3):115762. PubMed ID: 35988351. Abstract: The test performance and potential clinical utility of the ePlex® BCID Gram-Positive (GP) Panel was evaluated relative to MALDI-TOF mass spectrometry on bacterial isolates and traditional antimicrobial susceptibility testing. All GP bacteria (n = 100) in the study were represented on the panel including 50 common skin contaminants, and 7/7 coinfections. The positive percent agreement (PPA) was 97/97 with 2 false positives. Detection of vanA yielded a PPA of 4/4 and NPA of 9/9. mecA gene detection exhibited a PPA of 14/14 and NPA of 14/14 for S. aureus and a PPA of 31/32(97%) and NPA of 16/16 for CNS with 1 false negative. Chart reviews (n = 80) identified a mean 24.4h faster time to organism identification, 53.4h earlier optimization in 15(18.8%) patients based on AMR gene detection, 29.2h earlier optimization for 8(10%) patients infected with organisms, such as streptococci, with very low resistance rates, and 42.9h earlier discontinuation of antimicrobials for 14(17.5%) patients with contaminant cultures.[Abstract] [Full Text] [Related] [New Search]