These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The Effects of Filtering PPG Signal on Pulse Arrival Time-Systolic Blood Pressure Correlation. Author: Wang W, Marefat F, Mohseni P, Kilgore K, Najafizadeh L. Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2022 Jul; 2022():674-677. PubMed ID: 36086297. Abstract: Pulse arrival time (PAT), evaluated from electro-cardiogram (ECG) and photoplethysmogram (PPG) signals, has been widely used for cuff-less blood pressure (BP) estimation due to its high correlation with BP. However, the question of whether filtering the PPG signal impacts the extracted PAT values and consequently, the correlation between PAT and BP, has not been investigated before. In this paper, using data from 18 subjects, changes in the PAT values, and in the subject-specific PAT-systolic BP (SBP) correlation caused by filtering the PPG signal with variable cutoff frequencies in the range of 2 to 15 Hz are studied. For PAT extraction, three PPG characteristic points (foot, maximum slope and systolic peak) are considered. Results show that differences in the cutoff frequency can shift the PAT values and introduce a worst-case error of over 8.2 mmHg for SBP estimation, indicating that PPG signal filter settings can impact PAT-based BP estimations. Our study suggests that extracting the PAT from the maximum slope point of PPG signal filtered at 10 Hz provides the most stable correlation with SBP.[Abstract] [Full Text] [Related] [New Search]