These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Compatibility levels between blood gas analysis and central laboratory hemoglobin and electrolyte tests in pediatric patients: A single-center experience. Author: Konuksever D, Yucel SP, Bölük O, Kılıç BO, Taşar MA. Journal: Paediatr Anaesth; 2023 Feb; 33(2):107-113. PubMed ID: 36178754. Abstract: INTRODUCTION: We aimed to evaluate the interchangeability of sodium, potassium, hemoglobin, and hematocrit measurement between the blood gas analyzers and laboratory automatic analyzers results. METHODS: This was a retrospective cross-sectional study. The results of 1927 paired samples analyzed simultaneously with the blood gas analyzer and the laboratory automatic analyzer were compared. The Bland-Altman and Cohen's kappa statistic detected the agreement between the two analyses. RESULTS: The limits of agreement (±1.96 standard deviation of the mean difference) were -11.1 to 20.3 for sodium, -1.9 to 0.5 for potassium, -16.1 to 12.9 for hematocrit, and -5.0 to 4.0 for hemoglobin. Agreement between the two analyses was not acceptable within the defined clinically acceptable limits. In addition, none of the kappa values were higher than 0.60, which highlights the lack of agreement between the two analyzers. CONCLUSION: The blood gas analyzers and laboratory automatic analyzers results cannot be used interchangeably.[Abstract] [Full Text] [Related] [New Search]