These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Simultaneous analysis of acylcarnitines using MALDI-TOF mass spectrometry based on a parylene matrix chip. Author: Noh JY, Kim MJ, Yun TG, Kang MJ, Pyun JC. Journal: Analyst; 2022 Nov 07; 147(22):5044-5053. PubMed ID: 36196954. Abstract: Short and medium chain acylcarnitines have been used for the diagnosis of various fatty acid oxidation and organic acid disorders. This report presents a multiplex and quantitative analysis of acylcarnitines using MALDI-TOF MS based on a parylene matrix chip. The parylene matrix chip was fabricated by the deposition of a nanoporous film of parylene on an organic matrix array, which reduced the number of mass peaks from the organic matrix in the low m/z range. Quantitative analysis was possible using the parylene matrix chip because of the formation of nano-sized sample crystals on the nanoporous parylene film. Seven acylcarnitines were quantitatively analyzed using the chip; the method detection range included the cut-off values for metabolic disorders. The seven acylcarnitines of different concentrations were simultaneously detected using the parylene matrix chip and the interference from the mixed carnitines was estimated. Real L-carnitine (C0) samples were analyzed using serial dilution, and the recoveries were calculated by comparisons with a standard curve.[Abstract] [Full Text] [Related] [New Search]