  • Title: Effects of time delay and blood storage methods on analysis of canine venous blood samples with an Element point-of-care analyzer.
    Author: Rincon AA, Wurlod VA, Liu CC, Smith MR.
    Journal: Vet Clin Pathol; 2023 Mar; 52(1):22-29. PubMed ID: 36573472.
    Abstract:
    BACKGROUND: Manufacturers of point-of-care (POC) analyzers recommend immediate processing and anaerobic collection of blood samples. However, it is not uncommon for clinical scenarios to result in delayed sample processing or room air exposure that could impact the test results. OBJECTIVE: To investigate the effect of time delay and sample storage method on key POC analytes in canine venous blood samples processed with an Element POC analyzer. METHODS: Blood gas analysis was performed on venous blood samples at times 0 (T0), 15, 30, and 60 minutes after sampling using three different storage methods: preheparinized plastic syringes and two different lithium heparin tubes. To determine clinical relevance, results were compared with allowable total error of the respective parameter. Significance was set at P < 0.05. RESULTS: Significant differences between the three storage methods at baseline were found for partial pressure of carbon dioxide (PCO2 ), partial pressure of oxygen (PO2 ), base excess, and total hemoglobin. No significant differences up to T60 were found within collection methods for actual bicarbonate (HCO3 - ), base excess, sodium, potassium, chloride, ionized calcium (iCa), glucose, and BUN. Significant differences within collection methods were found after T0 for creatinine, after 15 minutes for lactate, and after 30 minutes for pH and hematocrit. No significant differences were found for PO2 in samples stored in preheparinized plastic syringes at any time point. CONCLUSIONS: These results suggest that HCO3 - , sodium, potassium, chloride, iCa, glucose, and BUN are comparable within the three storage methods for up to 60 minutes after sampling without resulting in clinically relevant changes.
    [Abstract] [Full Text] [Related] [New Search]