These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Errors in longitudinal measurements of bone mineral: effect of source strength in single and dual photon absorptiometry. Author: Dunn WL, Kan SH, Wahner HW. Journal: J Nucl Med; 1987 Nov; 28(11):1751-7. PubMed ID: 3668666. Abstract: The effect of changing strength during the useful life of a radiation source was evaluated in studies performed on four dual photon (DPA) and two single photon (SPA) bone absorptiometry instruments. Two DPA units and one SPA unit did not show any systematic dependence of measured bone mineral content or bone mineral areal density (BMD) on source activity when evaluated over an entire source life. One DPA and one SPA instrument, however, showed significant time trends associated with source activity. The fourth DPA instrument had a significant linear decrease in BMD over a source life in the automatic mode but performed better in the manual mode.[Abstract] [Full Text] [Related] [New Search]