  • Title: Retrospective analysis of intra-patient laboratory variation demonstrates that the BD Vacutainer® Barricor™ blood collection tube reduces troponin variation.
    Author: Cembrowski G, Qiu Y, Sherazi A, Shea JL.
    Journal: Clin Biochem; 2023 Apr; 114():24-29. PubMed ID: 36706798.
    Abstract:
    OBJECTIVE: The BD Vacutainer® Barricor™ plasma blood collection tube uses a mechanical separator during centrifugation to separate plasma from the cellular elements of blood. Compared to use of plasma separator tubes (PST™) with gel, Barricor™ produces a cleaner sample with less residual cellular content. We sought to determine if Barricor™ reduces pre-analytical error compared to PST™. DESIGN & METHODS: We used a model previously published that utilizes serial differences between intra-patient measurements transformed into a Taylor series of variation vs time with the y-intercept equal to the sum of short-term analytic variation, preanalytic variation and biologic variation. The intra-patient variation of chloride, sodium, potassium, and troponin-T (hs-TnT) obtained from the Emergency Department of a large tertiary care center sampled with PST™ (May 2015-April 2018, n = 59,762 specimens) or Barricor™ (May 2018-May 2021, n = 61,512 specimens) was evaluated. All specimens were analyzed on either Roche Modular or Cobas® instruments. For each analyte, pairs of intra-patient results were tabulated and separated by 1 h intervals. The average between-pair variations were then regressed against time. We also determined the number of intra-patient outliers using the reference change value for each analyte. RESULTS: The Barricor™ hs-TnT y-intercept (-0.0132) was significantly lower than the PST™ intercept (0.9109; p = 0.022). This was also true for chloride (y-intercept = 1.0067 in Barricor™ and 1.3431 in PST™, p = 0.037). The percentage of hs-TnT outliers was significantly lower in Barricor™ (8.32 %) vs PST™ (12.2 %; p < 0.001). CONCLUSION: The analytical and biological variations are assumed to be steady over the study periods; we ascribe the difference in the y-intercept to the preanalytical effect of the Barricor™ tube reducing platelets and other cellular debris.
    [Abstract] [Full Text] [Related] [New Search]