These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Audiological outcome after stapes surgery in relation to prosthesis type. Author: Handke V, Agha-Mir-Salim P, James P, Müller A. Journal: Eur Arch Otorhinolaryngol; 2023 Jul; 280(7):3171-3176. PubMed ID: 36707432. Abstract: PURPOSE: Different techniques are used to fix crimp and CliP® Piston stapes prostheses to the long process of the incus (LPI). The CliP® Piston provides a stiff connection in contrast to the static bended loop of the crimp prosthesis, which imitates the physiological incudostapedial joint (ISJ) and thereby potentially leads to different hearing outcome. METHODS: In a retrospective single-center study of German-speaking one hundred and ninety patients who underwent stapes surgery CliP® Piston or crimp prostheses between the years of 2016 and 2019 by the same surgeon and in the same setting. Pre- and postoperative bone- (BC) and air-conduction (AC) pure-tone thresholds, pre- and postoperative air-bone gap (ABG) for 0.5, 1, 1.5, 2, 3, 4 kHz and the surgery time were examined. RESULTS: The postoperative bone conduction thresholds were significantly lower in the frequencies between 0.5 and 3 kHz and the mean ABG was < 10 dB in most cases independent of the prosthesis used. Crimp prosthesis showed a significantly better closure of the ABG at 0.5 kHz. CONCLUSIONS: The audiological outcome after stapes surgery is dependent on the type of prosthesis used, as reflected by the frequency-specific air-bone gap. The better ABG closure with the crimp prosthesis might be the result of the connection to the LPI imitating the physiological ISJ. The crimp prosthesis may be the better choice if use of hearing aids is expected postoperatively.[Abstract] [Full Text] [Related] [New Search]