  • Title: Development of loop-mediated isothermal amplification (LAMP) assays using five primers reduces the false-positive rate in COVID-19 diagnosis.
    Author: Alhamid G, Tombuloglu H, Al-Suhaimi E.
    Journal: Sci Rep; 2023 Mar 28; 13(1):5066. PubMed ID: 36977756.
    Abstract:
    The reverse-transcription loop-mediated isothermal amplification (RT-LAMP) is a cheaper and faster testing alternative for detecting SARS-CoV-2. However, a high false-positive rate due to misamplification is one of the major limitations. To overcome misamplifications, we developed colorimetric and fluorometric RT-LAMP assays using five LAMP primers, instead of six. The gold-standard RT-PCR technique verified the assays' performance. Compared to other primer sets with six primers (N, S, and RdRp), the E-ID1 primer set, including five primers, performed superbly on both colorimetric and fluorometric assays. The sensitivity of colorimetric and fluorometric assays was 89.5% and 92.2%, respectively, with a limit of detection of 20 copies/µL. The colorimetric RT-LAMP had a specificity of 97.2% and an accuracy of 94.5%, while the fluorometric RT-LAMP obtained 99% and 96.7%, respectively. No misamplification was evident even after 120 min, which is crucial for the success of this technique. These findings are important to support the use of RT-LAMP in the healthcare systems in fighting COVID-19.