These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Simple liquid-chromatographic determination of urinary coproporphyrin in workers exposed to lead.
    Author: Tomokuni K, Hirai Y.
    Journal: Clin Chem; 1986 May; 32(5):872-3. PubMed ID: 3698281.
    Abstract:
    We developed a simple method for the determination of urinary coproporphyrin (CP) in lead workers, using a "high-performance" liquid chromatograph (HPLC) equipped with a fluorescence detector. The detection limit of urinary CP in this method was 5 micrograms per liter of urine. The working linear range of urinary CP concentration was 5 to 1500 micrograms/L. In 41 lead-exposed workers, the urinary CP values obtained by the present HPLC method were well correlated with those obtained by a conventional spectrophotometric method (r = 0.94). The present method is useful for screening workers exposed to lead.
    [Abstract] [Full Text] [Related] [New Search]