These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Compression-associated injuries using CLOVER3000 device in non-survivor patients of OHCA: A retrospective cohort study. Author: Hayashi M, Tanizaki S, Nishida N, Mizuno H, Kano K, Tanaka J, Azuma H, Sera M, Nagai H, Maeda S. Journal: Am J Emerg Med; 2023 Jun; 68():127-131. PubMed ID: 36996591. Abstract: AIM: The incidence of compression-associated injuries from using the CLOVER3000, a new mechanical cardiopulmonary resuscitation (CPR) device, is not well studied in the context of out-of-hospital cardiac arrest (OHCA). Thus, we aimed to compare compression-associated injuries between CLOVER3000 and manual CPR. METHODS: This single-center, retrospective, cohort study used data from the medical records of a tertiary care center in Japan between April 2019 and August 2022. We included adult non-survivor patients with non-traumatic OHCA who were transported by emergency medical services and underwent post-mortem computed tomography. Compression-associated injuries were tested using logistic regression models adjusting for age, sex, bystander CPR performance, and CPR duration. RESULTS: A total of 189 patients (CLOVER3000, 42.3%; manual CPR, 57.7%) were included in the analysis. The overall incidence of compression-associated injuries was similar between the two groups (92.5% vs. 94.54%; adjusted odds ratio (AOR), 0.62 [95% confidence interval (CI), 0.06-1.44]). The most common injury was anterolateral rib fractures with a similar incidence between the two groups (88.7% vs. 88.9%; AOR, 1.03 [95% CI, 0.38 to 2.78]). The second most common injury was sternal fracture in both groups (53.1% vs. 56.7%; AOR, 0.68 [95% CI, 0.36-1.30]). The incidence rates of other injuries were not statistically different between the both groups. CONCLUSION: We observed a similar overall incidence of compression-associated injuries between the CLOVER3000 and manual CPR groups on small sample size.[Abstract] [Full Text] [Related] [New Search]