  • Title: Haemolysis as an interference factor in clinical chemistry.
    Author: Sonntag O.
    Journal: J Clin Chem Clin Biochem; 1986 Feb; 24(2):127-39. PubMed ID: 3711797.
    Abstract:
    Using fully mechanized analytical equipment, interference by haemolysis in the determination of 26 clinical chemical parameters was determined quantitatively by adding haemolysate to serum. Haemoglobin concentrations up to 6.6 g/l caused essentially no interference in the following determinations: albumin (immuno-nephelometric), alpha-amylase, calcium, chloride, cholesterol, cholinesterase, creatinine, iron, glucose, glutamate dehydrogenase, uric acid, urea, sodium, inorganic phosphate, total protein, transferrin and triglycerides. In the presence of haemoglobin, erroneously high values were found for: lactate dehydrogenase (haemoglobin higher than 0.2 g/l), aspartate aminotransferase, potassium and acid phosphate (haemoglobin higher than 1.5 g/l), creatine kinase (haemoglobin higher than 2.5 g/l) and alanine aminotransferase (haemoglobin higher than 3.4 g/l). Erroneously low values were found for bilirubin (haemoglobin higher than 0.8 g/l), alkaline phosphatase and albumin (by electrophoresis) (haemoglobin higher than 1.5 g/l) and gamma-glutamyltransferase (haemoglobin higher than 3.0 g/l).
    [Abstract] [Full Text] [Related] [New Search]