These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Repeatability and reproducibility of a new fully automatic measurement optical low coherence reflectometry biometer and agreement with swept-source optical coherence tomography-based biometer.
    Author: Yu J, Zhao G, Lei CS, Wan T, Ning R, Xing W, Ma X, Pan H, Savini G, Schiano-Lomoriello D, Zhou X, Huang J.
    Journal: Br J Ophthalmol; 2024 May 21; 108(5):673-678. PubMed ID: 37142332.
    Abstract:
    AIMS: To assess the repeatability and reproducibility of the ocular measurements obtained with the Suoer SW-9000 μm Plus, a new fully automatic biometer based on optical low coherence reflectometry (OLCR) biometer, and to compare them to those obtained by a swept-source optical coherence tomography (SS-OCT)-based biometer. METHODS: This prospective study consisted of 115 eyes of 115 healthy subjects. The measurements were taken by the two optical biometers in random order. The measured parameters were axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), mean keratometry (Km), lens thickness (LT) and corneal diameter (CD). To evaluate the intraobserver repeatability and interobserver reproducibility, the within-subject SD, test-retest variability, coefficient of variation (CoV) and intraclass correlation coefficient (ICC) were adopted. The Bland-Altman plot was drawn to assess the agreement. RESULTS: The repeatability and reproducibility of all parameters for the new device were excellent (ICC>0.960 and CoV<0.71%). The Bland-Altman plots showed high agreement between the OLCR-based and SS-OCT-based devices for AL, CCT, AQD, ACD, Km and LT, with narrow 95% limit of agreements (LoAs) (-0.08 mm to 0.06 mm, -15.91 µm to -1.01 µm, -0.09 mm to 0.09 mm, -0.09 mm to 0.08 mm, -0.47 D to 0.35 D, -0.05 mm to 0.16 mm, respectively) and moderate agreement for CD (95% LoA: -0.67 mm to -0.01 mm). CONCLUSIONS: The new Suoer SW-9000 μm Plus biometer showed excellent repeatability and reproducibility. All the parameters obtained by this biometer were similar to those measured by SS-OCT-based biometer.
    [Abstract] [Full Text] [Related] [New Search]