Title: Rapid cation-exchange high-performance liquid chromatographic procedure for the separation and quantitation of hemoglobins S, C, and O Arab in cord blood samples. Author: Wilson JB, Wrightstone RN, Huisman TH. Journal: J Lab Clin Med; 1986 Aug; 108(2):138-41. PubMed ID: 3734587. Abstract: A rapid method for separating and quantitating hemoglobin (Hb) variants in cord blood samples using cation high-performance liquid chromatography (HPLC) is described. The procedure is a modification of a previously published method, and uses a weak cation-exchange Brownlee-3CM column and Bis-Tris-KCN-Na acetate developers. A chromatogram can be completed in 10 minutes. The slow-moving variants, Hbs S, C, and O Arab, can be completely separated from each other and are identified by their elution times relative to Hb A. Hb E elutes as a shoulder on the descending side of Hb A, which is characteristic for this variant in this procedure. Differentiation between heterozygous, homozygous, and Hb X-beta+-thalassemia conditions is easily made.[Abstract] [Full Text] [Related] [New Search]