These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The effects of topping-off instrumentation on biomechanics of sacroiliac joint after lumbosacral fusion.
    Author: Fan W, Zhang C, Wang QD, Guo LX, Zhang M.
    Journal: Comput Biol Med; 2023 Sep; 164():107357. PubMed ID: 37586205.
    Abstract:
    BACKGROUND: Lumbar/lumbosacral fusion supplemented with topping-off devices has been proposed with the aim of avoiding adjacent segment degeneration proximal to the fusion construct. However, it remains unclear how the biomechanics of the sacroiliac joint (SIJ) are altered after topping-off surgery. The objective of this study was to investigate the biomechanical effects of topping-off instrumentation on SIJ after lumbosacral fusion. METHODS: The validated finite element model of an intact lumbar spine-pelvis segment was modified to simulate L5-S1 interbody fusion fixed with a pedicle screw system. An interspinous spacer, Device for Intervertebral Assisted Motion (DIAM), was used as a topping-off device and placed between interspinous processes of the L4 and L5 segments. Range of motion (ROM), von-Mises stress distribution, and ligament strain at SIJ were compared between fusion (without DIAM) and topping-off (fusion with DIAM) models under moments of four physiological motions. RESULTS: ROM at the left and right SIJs in the topping-off model was higher by 26.9% and 27.5% in flexion, 16.8% and 16.1% in extension, 18.8% and 15.8% in lateral bending, and 3.7% and 7.4% in axial rotation, respectively, compared to those in the fusion model. The predicted stress and strain data showed that under all physiological loads, the topping-off model exhibited higher stress and ligament strain at the SIJs than the fusion model. CONCLUSIONS: Motion, stress, and ligament strain at SIJ increase when supplementing lumbosacral fusion with topping-off devices, suggesting that topping-off surgery may be associated with higher risks of SIJ degeneration and pain than fusion alone.
    [Abstract] [Full Text] [Related] [New Search]