These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Disc embolization of Björk-Shiley mitral prosthesis without strut fracture. Author: McGeehin WH, MacVaugh H, Donahoo JS, Lechman MJ. Journal: J Thorac Cardiovasc Surg; 1986 Oct; 92(4):795-6. PubMed ID: 3762210. Abstract: Strut fracture with subsequent embolization of the Björk-Shiley mitral prosthesis is a rare but recognized complication. A case of disc embolization without strut fracture was managed successfully by early surgical intervention. The preoperative chest x-ray film demonstrating apparently intact major and minor struts underscores the importance of a high index of suspicion to make an early diagnosis.[Abstract] [Full Text] [Related] [New Search]