These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Radiation of the fiberoptic illuminator of an operating microscope.
    Author: Aarnisalo E.
    Journal: Acta Ophthalmol (Copenh); 1986 Aug; 64(4):397-400. PubMed ID: 3776503.
    Abstract:
    Relative radiation spectrum of a fiberoptic illuminator of an operating microscope (Zeiss Opmi 6) was measured with a spectrophotometer within the range of 350 to 1100 nm. Total irradiance at the corneal level was measured with a radiometer within the range of 250 to 4000 nm. Illumination was measured with a standard illuminometer. Two selective filters were used to optimally reduce the total radiation. A Schott filter GG 420 was used to cut off ultraviolet radiation and a Schott thermal filter KG 1 to reduce intrared radiation. Measured without filters the energy of radiation of wavelengths shorter than 400 nm was only 0.05% of the total energy of radiation, 78.5% was within the range of 400 to 780 nm, and 21.4% was produced by radiation of wavelengths longer than 780 nm. Corresponding values measured with selective filters were 0.003, 95.4 and 4.6. Irradiance at the corneal level was 7.7 mW/cm2 without filters and 3.8 mW/cm2 with selective filters, and the corresponding illumination levels were 21,300 and 16,500 lux. Using selective filters the total irradiance was effectively reduced causing only a minor reduction of the illumination level and a small just noticeable alteration of the colour of the operating field.
    [Abstract] [Full Text] [Related] [New Search]