These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Anaesthetic scavenger evaluation in the laboratory. Author: Wetterhahn NS, Cleaton-Jones PE, Moyes DG, Fatti LP. Journal: Eur J Anaesthesiol; 1986 Mar; 3(2):95-102. PubMed ID: 3780689. Abstract: A method of testing scavenging devices in isolation in a test box was developed. The rate of pollution in the box was measured with varying ventilating and suction conditions which allowed a leak index to be calculated so that the efficacy of commercially available scavengers could be compared to the system without a scavenger, and to each other. The BOC scavenger tested reduced pollution but was less efficient at low suction rates. For this scavenger the suction rate should be greater than 20 litre min-1.[Abstract] [Full Text] [Related] [New Search]