These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Diagnostic sample transport via pneumatic tube systems: data logger and their algorithms are sensitive to transport effects.
    Author: Ninnemann J, Zylla S, Streichert T, Otto B, Haenel M, Nauck M, Petersmann A.
    Journal: Clin Chem Lab Med; 2024 Mar 25; 62(4):657-663. PubMed ID: 37833063.
    Abstract:
    OBJECTIVES: Many hospitals use pneumatic tube systems (PTS) for transport of diagnostic samples. Continuous monitoring of PTS and evaluation prior to clinical use is recommended. Data loggers with specifically developed algorithms have been suggested as an additional tool in PTS evaluation. We compared two different data loggers. METHODS: Transport types - courier, conventional (cPTS) and innovative PTS (iPTS) - were monitored using two data loggers (MSR145® logger, CiK Solutions GmbH, Karlsruhe, Germany, and a prototype developed at the University Medicine Greifswald). Data loggers differ in algorithm, recording frequencies and limit of acceleration detection. Samples from apparently healthy volunteers were split among the transport types and results for 37 laboratory measurands were compared. RESULTS: For each logger specific arbitrary units were calculated. Area-under-the-curve (AUC)-values (MSR145®) were lowest for courier and highest for iPTS and increased with increasing recording frequencies. Stress (St)-values (prototype logger) were obtained in kmsu (1,000*mechanical stress unit) and were highest for iPTS as well. Statistical differences between laboratory measurement results of transport types were observed for three measurands sensitive for hemolysis. CONCLUSIONS: The statistical, but not clinical, differences in the results for hemolysis sensitive measurands may be regarded as an early sign of preanalytical impairment. Both data loggers record this important interval of beginning mechanical stress with a high resolution indicating their potential to facilitate early detection of preanalytical impairment. Further studies should identify suitable recording frequencies. Currently, evaluation and monitoring of diagnostic sample transport should not only rely on data loggers but also include diagnostic samples.
    [Abstract] [Full Text] [Related] [New Search]