These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Large animal ventilator-integrated volumetric capnography generates clinically acceptable values of physiologic dead space in anesthetized healthy adult horses.
    Author: Frampton A, Floriano D, Simpson K, Hopster K.
    Journal: Am J Vet Res; 2024 Jan 01; 85(1):. PubMed ID: 37857347.
    Abstract:
    OBJECTIVE: To evaluate the agreement between the Tafonius large animal ventilator-integrated volumetric capnography (vCap) software and the Respironics NICO noninvasive cardiac output monitor reference system. ANIMALS: Data were collected from 56 healthy adult horses undergoing general anesthesia. METHODS: Animals were placed under general anesthesia and connected to the Tafonius large animal ventilator circle system. A flow partitioning device with CO2 and flow sensors was utilized to couple the endotracheal tube to the NICO monitor. Tafonius CO2 and flow sensors are incorporated into the Y-piece of the breathing circuit. Arterial blood samples were collected to determine the partial pressure of arterial carbon dioxide (PaCO2) immediately before data collection. The PaCO2 was input into the Tafonius and NICO monitor, and dead space ventilation (%Vd), end-tidal CO2 partial pressure (ETco2), mixed-expired CO2 partial pressure (Peco2), and expired tidal volume (Vt) were calculated over a single breath. Multiple measurements were completed for each patient, with a total of 200 paired data points collected for analysis. Data were assessed for normality, and Bland-Altman analysis was performed. Bias and 95% limits of agreement were calculated. RESULTS: The limits of agreement for %Vd of the ventilator-derived measurements fell within ± 10% of the NICO monitor reference method. CLINICAL RELEVANCE: Our results indicate that, when compared to the NICO monitor method, the Tafonius-integrated vCap software provides clinically acceptable values of Peco2, Vt, and %Vd in healthy adult horses.
    [Abstract] [Full Text] [Related] [New Search]