  • Title: Effect of intravenous vitamin C on arterial blood gas analyser and Accu-Chek point-of-care glucose monitoring in critically ill patients.
    Author: Rosengrave PC, Wohlrab C, Spencer E, Williman J, Shaw G, Carr AC.
    Journal: Crit Care Resusc; 2022 Jun 06; 24(2):175-182. PubMed ID: 38045598.
    Abstract:
    Background: Intravenous vitamin C is known to interfere with some point-of-care blood glucose meters. We aimed to determine the concentrations at which ascorbate interferes with glucose concentrations measured using a point-of-care blood glucose meter. We also compared the point-of-care meter and an arterial blood gas (ABG) analyser in the intensive care unit with laboratory glucose monitoring in septic patients receiving intravenous vitamin C infusions. Methods: Blood samples containing normal, depleted and supplemented glucose and increasing concentrations of ascorbate (0.1-1.0 mmol/L) were tested using an Accu-Chek Inform II (Roche Diagnostics, USA) glucometer. For the in vivo study, 41 individual blood samples were drawn daily from septic patients (n = 16) receiving infusions of 25 mg/kg of vitamin C every 6 hours. The glucose values of matched blood samples were assessed using Accu-Chek, ABG and laboratory glucose methods. Results: For every 1 mmol/L of ascorbate added, the glucose concentration measured by the point-of-care monitor increased by 1.4 mmol/L (95% CI, 1.0-1.8; P < 0.001). Analysis of matched blood samples collected following intravenous vitamin C infusion indicated that 98% of the ABG and 83% of the Accu-Chek values met the International Organization for Standardization (ISO) 15197:2013 accuracy criteria. One patient had severe renal impairment, which contributed to elevated plasma vitamin C concentrations (median, 0.95 mmol/L; range, 0.64-1.10 mmol/L), resulting in elevated Accu-Chek readings and presenting a moderate clinical risk for the highest value. Conclusions: Vitamin C concentrations < 0.8 mmol/L do not interfere with point-of-care glucose monitoring. Intravenous vitamin C infusion of 25 mg/kg every 6 hours does not interfere with point-of-care glucose monitoring unless the patient has renal impairment, in which case laboratory glucose tests should be used.
    [Abstract] [Full Text] [Related] [New Search]