These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of thallium in urine with Zeeman effect graphite furnace atomic absorption.
    Author: Paschal DC, Bailey GG.
    Journal: J Anal Toxicol; 1986; 10(6):252-4. PubMed ID: 3807326.
    Abstract:
    Thallium is determined in urine with a simple dilution as the only sample pretreatment. Zeeman effect background correction provides adequate accommodation for non-atomic absorption such that sufficient diluted urine can be introduced into the furnace to achieve a detection limit of 0.5 ng/mL. Precision and accuracy were evaluated with Environmental Protection Agency water samples and in vitro spiked urine pools; precision of 15% at the 6 ng/mL level was observed. Diluted urine was used in calibration, with only 4 to 5% difference in slope observed within run among randomly selected urine specimens.
    [Abstract] [Full Text] [Related] [New Search]