These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Differential white cell counting on the Coulter Counter Model S Plus IV (three population) and the Technicon H6000; a comparison by simple and multiple regression.
    Author: Clarke PT, Henthorn JS, England JM.
    Journal: Clin Lab Haematol; 1985; 7(4):335-51. PubMed ID: 3830531.
    Abstract:
    Simple and multiple regression analyses have been used to see which of the differential white count measurements on the Technicon H6000 could be used to predict the differential white count measurements on the Coulter Counter Model S Plus IV (three population). The S Plus IV lymphocytes are predicted by the Technicon H6000 lymphocytes plus 39% of monocytes. The S Plus IV granulocytes are predicted by the Technicon H6000 neutrophils plus most of eosinophils, plus 43% of large unstained cells. The S Plus IV mononuclear cells are predicted by 5% of the Technicon H6000 lymphocytes plus 41% of monocytes, plus basophils, plus most of the large unstained cells plus 31% of eosinophils.
    [Abstract] [Full Text] [Related] [New Search]