These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Most recently developed polyester polymer alloy dialyzer: A new medium cut-off membrane? Author: Maduell F, Rodríguez-Espinosa D, Rodas LM, Gómez M, Arias-Guillén M, Fontseré N, Vera M, Del Carmen Salgado M, Rico N, Broseta JJ. Journal: Artif Organs; 2024 Jul; 48(7):753-762. PubMed ID: 38404240. Abstract: BACKGROUND: New versions of the polyester polymer alloy (PEPA) membrane have appeared over the years, with increases in both the pore size and the amount of polyvinylpyrrolidone (PVP) to optimize hydrophilicity performance. This study aimed to assess the efficacy of the most recently developed PEPA dialyzer, the FDY series, in hemodialysis (HD) modality in terms of uremic toxin removal and albumin loss and to compare it with that of several high-flux dialyzers currently used in HD and post-dilution hemodiafiltration (HDF) treatments. METHODS: A prospective study was carried out in 21 patients. All patients underwent six dialysis sessions with the same routine dialysis parameters; only the dialyzer and/or the dialysis modality varied: FX80 in HD, FDY 180 in HD, Clearum HS17 in HDF, Elisio 19H in HDF, Vitapes 180 in HDF, and FX80 in post-dilution HDF. The reduction ratios (RR) of urea, creatinine, ß2-microglobulin, myoglobin, κFLC, prolactin, α1-microglobulin, α1-acid glycoprotein, λFLC, and albumin were compared intraindividually. Dialysate albumin loss was also measured. RESULTS: Both membranes FDY and FX80 are high-flux dialyzers and are applied here in high-flux HD. The average RR of β2-microglobulin was slightly lower in the two HD treatments than in the HDF treatments. Comparison of dialysis treatments revealed that the PEPA FDY dialyzer in the HD modality was more effective than the FX80 dialyzer in high-flux HD and was as effective as post-dilution HDF, especially in terms of myoglobin, κFLC, prolactin, α1-microglobulin, and λFLC RRs. The FDY treatments obtained similar albumin RR in blood and slightly higher dialysate albumin loss, although the values were clinically acceptable. CONCLUSIONS: The most recently developed PEPA dialyzers in the HD modality were as effective as all treatments in the HDF modality and were clearly superior to high-flux helixone HD treatment. These results confirm that this dialyzer should be categorized within the medium cut-off (MCO) membrane classification.[Abstract] [Full Text] [Related] [New Search]