Search MEDLINE/PubMed
Title: Cardiac implantable electronic device-induced tricuspid regurgitation: Implications and management. Author: Murphy SP, Lew J, Yucel E, Singh J, Mela T. Journal: J Cardiovasc Electrophysiol; 2024 May; 35(5):1017-1025. PubMed ID: 38501386. Abstract: Tricuspid regurgitation (TR) secondary to cardiac implantable electronic devices (CIEDs) has been well documented and is associated with worse cardiovascular outcomes. A variety of mechanisms have been proposed including lead-induced mechanical disruption of the tricuspid valvular or subvalvular apparatus and pacing-induced electrical dyssynchrony. Patient characteristics such as age, sex, baseline atrial fibrillation, and pre-existing TR have not been consistent predictors of CIED-induced TR. While two-dimensional echocardiography is helpful in assessing the severity of TR, three-dimensional echocardiography has significantly improved accuracy in identifying the etiology of TR and whether lead position contributes to TR. Three-dimensional echocardiography may therefore play a future role in optimizing lead positioning during implant to reduce the risk of CIED-induced TR. Optimal lead management strategies in addition to percutaneous interventions and surgery in alleviating TR are very important.[Abstract] [Full Text] [Related] [New Search]