  • Title: Rapid quantitative, specific measurement of pancreatic amylase in serum with use of a monoclonal antibody.
    Author: Mifflin TE, Benjamin DC, Bruns DE.
    Journal: Clin Chem; 1985 Aug; 31(8):1283-8. PubMed ID: 3874721.
    Abstract:
    In this rapid quantitative assay for pancreatic alpha-amylase (EC 3.2.1.1) in serum, we precipitate salivary amylases by 10-min incubation with monoclonal anti-salivary amylase antibody immobilized on particles of polyvinylidene fluoride. We then centrifuge the serum mixture and measure the pancreatic amylase activity remaining in the supernate by a kinetic method. The assay requires 50 microL of serum and the standard curve is linear to at least 1300 U of pancreatic amylase per liter of serum. CVs were 1.3% within-run, 6-8% day-to-day. Apparent analytical recovery of pancreatic amylase activity added to serum was 101% +/- 2%. Addition of purified salivary amylase, 356 U/L, to sera gave a value for apparent pancreatic amylase of less than 4 U/L, or 1% of the added salivary amylase activity. This assay correlated well with an electrophoretic method (slope, 0.97-0.99; intercept, 0.5 to -4 U/L; correlation coefficient, 0.946-0.990; and standard error of the estimate 3-5 U/L). Estimated normal reference intervals with maltotetraose as substrate were: total amylase, 39-118 U/L; pancreatic amylase, 11-50 U/L; and salivary amylase, 18-79 U/L.