These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Validation of smartwatch electrocardiogram intervals in children compared to standard 12 lead electrocardiograms. Author: Ernstsson J, Svensson B, Liuba P, Weismann CG. Journal: Eur J Pediatr; 2024 Sep; 183(9):3915-3923. PubMed ID: 38918230. Abstract: UNLABELLED: Lay people are now able to obtain one-lead electrocardiograms (ECG) using smartwatches, which facilitates documentation of arrhythmias. The accuracy of smartwatch derived ECG intervals has not been validated in children though. Home-based monitoring of ECG intervals using a smartwatch could improve monitoring of children, e.g. when taking QTc prolonging medications. The aim of this study was to validate the ECG intervals measured by smartwatch in comparison to standard 12-lead ECGs in children and adolescents. Prospective study of children (age 5-17 years) at the outpatient clinic of a national pediatric heart center. Patients underwent a smartwatch ECG (ScanWatch, Withings) and a simultaneous standard 12-lead ECG. ECG intervals were measured both automatically and manually from the smartwatch ECG and the 12-lead ECG. Intraclass correlation coefficients and Bland-Altman plots were performed. 100 patients (54% male, median age 12.9 (IQR 8.7-15.6) were enrolled. The ICC calculated from the automated smartwatch and automated 12-lead ECG were excellent for heart rate (ICC 0.97, p < 0.001), good for the PR and QT intervals (ICC 0.86 and 0.8, p < 0.001), and moderate for the QRS duration and QTc interval (ICC 0.7 and 0.53, p < 0.001). When using manual measurements for the smartwatch ECG, validity was improved for the PR interval (ICC 0.93, p < 0.001), QRS duration (ICC 0.92, p < 0.001), QT (ICC 0.95, p < 0.001) and QTc interval (ICC 0.84, p < 0.001). CONCLUSION: Automated smartwatch intervals are most reliable measuring the heart rate. The automated smartwatch QTc intervals are less reliable, but this may be improved by manual measurements. WHAT IS KNOWN: In adults, smartwatch derived ECG intervals measured manually have previously been shown to be accurate, though agreement for automated QTc may be fair. WHAT IS NEW: In children, automated smartwatch QTc intervals are less reliable than RR, PR, QRS and uncorrected QT interval. Accuracy of the QTc can be improved by peroforming manual measurements.[Abstract] [Full Text] [Related] [New Search]