These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Development of an amperometric biosensor for the detection of Bacillus anthracis spores. Author: Ortatatli M, Saatci E, Kilic Z, Kenar L. Journal: Diagn Microbiol Infect Dis; 2024 Nov; 110(3):116473. PubMed ID: 39128207. Abstract: INTRODUCTION: Due to most likely use of Bacillus anthracis in biological terrorism agents, the rapid and sensitive detection of its spores is crucial in both taking prophylactic measures and proper treatment. This study aimed to develop an amperometric electrochemical immunosensor for the detection of B. anthracis spores. METHODS: A new amperometric biosensor was designed using a combination of magnetic beads and multiplex screen-printed electrodes. This method measures changes in current intensity resulting from oxidation and reduction in the working electrode directly to spore concentrations. RESULTS: A standard curve was formed to test the number of live spores between 2 × 102-2 × 104 spores/ml concentrations. LOD and LOQ values were found to be 92 and 272 spores/ml, respectively. No cross-reactions were seen for Bacillus subtilis, Bacillus cereus and Bacillus thuringiencis spores. CONCLUSIONS: It is shown that the designed Anthrax immunosensor has high sensitivity and selectivity with rapid detection results.[Abstract] [Full Text] [Related] [New Search]