These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Cardiac output determination: thermodilution versus a new computerized Fick method.
    Author: Carpenter JP, Nair S, Staw I.
    Journal: Crit Care Med; 1985 Jul; 13(7):576-9. PubMed ID: 3924481.
    Abstract:
    A computerized system is described for the on-line analysis of respiratory flow, complete analysis of inspired and expired gas, and computation of oxygen consumption and carbon dioxide production. The system has been compared to traditional Douglas bag determinations of these quantities. Fick cardiac output using computed oxygen consumption (VO2) was highly correlated (r = .92, p less than .001) with thermodilution cardiac output, determined simultaneously. Moreover, the computerized Fick method was much more reproducible.
    [Abstract] [Full Text] [Related] [New Search]