These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Automated platelet count in thrombocytopenic patients--a comparison of methods]. Author: Stibbe W, Weise M, Seidel D. Journal: J Clin Chem Clin Biochem; 1985 Jul; 23(7):399-404. PubMed ID: 3930654. Abstract: 157 EDTA blood samples from thrombocytopenic patients (platelets less than 100 X 10(9)/1) were measured and the platelet counts compared using the various blood cell counters of the central laboratory, University Clinic, Göttingen. Using the H 6000, Hemalog 8 (both Technicon) and ELT-8 (Ortho Instruments), product-moment correlation coefficients (r) ranging from 0.88 to 0.92 were obtained between the different counters. The correlation coefficients of the thrombocounter (Coulter) to the other instruments were only 0.84 and 0.85. Discrepant platelet counts (difference greater than 40 X 10(9)/1) were reinvestigated using the Neubauer chamber. From a total of 354 platelet measurements, falsely high concentrations were observed only in three cases and falsely low concentrations in seven cases.[Abstract] [Full Text] [Related] [New Search]