These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of oxytetracycline in premixes and veterinary products by liquid chromatography. Author: Chappell GS, Houglum JE, Kelley WN. Journal: J Assoc Off Anal Chem; 1986; 69(1):28-30. PubMed ID: 3949693. Abstract: A liquid chromatographic method for the assay of oxytetracycline in premixes and veterinary products is described. Premix samples are extracted with acidified methanol, diluted with mobile phase, and filtered before chromatography on a C-8, reverse phase column. The assay method separates oxytetracycline from epioxytetracycline, tetracycline, and chlortetracycline. Total elution time for oxytetracycline is less than 5 min at 1.5 mL/min. Five spiked premix samples each of 2 and 50 g/lb had a coefficient of variation of 3.5 and 4.5% and a mean recovery of 99 and 104%, respectively. The results of premixes and veterinary products assayed by this method compared closely with those of the same assayed by the official AOAC microbiological method.[Abstract] [Full Text] [Related] [New Search]