These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The accuracy of esophageal pressure measurements in convalescent and sick intubated infants. Author: Heaf DP, Turner H, Stocks J, Helms P. Journal: Pediatr Pulmonol; 1986; 2(1):5-8. PubMed ID: 3951892. Abstract: Assessment of the esophageal pressure technique for measuring dynamic lung compliance (Cdyn) showed that accurate results were obtained in only 5 of 15 studies of sick intubated infants, whereas they were obtained in 6 of 6 studies of convalescent nonintubated infants. In sick infants, inaccuracy was due to large variations in esophageal pressure change (delta Pes) at different esophageal depths and difficulties validating Pes using the occlusion test. Methods of assessing lung stiffness that depend on the measurement of esophageal pressure are unreliable and should not be used in sick intubated infants.[Abstract] [Full Text] [Related] [New Search]