These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Practical gamma camera quality control with a four-point phantom. Author: Lancaster JL, Kopp DT, Lasher JS, Blumhardt R. Journal: J Nucl Med; 1985 Mar; 26(3):300-7. PubMed ID: 3973741. Abstract: A gamma camera QC phantom for practical assessment of resolution (average FWHM), spatial calibration (pixels/mm), nonhomogeneity of spatial calibration, and average point-source sensitivity (cpm/muCi) was developed and evaluated. The phantom consists of four 57Co-point sources mounted on a Plexiglas base at corners of a square 10 cm on each side. Computer acquisition and processing are fully automated and require less than 1 min for point sources totaling 100 muCi (3.7 MBq) activity. The normal range of variability of measured QC parameters from a 12 wk evaluation period with four different gamma cameras (assumed to be operating normally) yielded coefficients of variation ranging from 0.3% for spatial calibration (pixels/mm) to 2.3% for sensitivity (cpm/muCi) assessments. From the normal range of variability a minimum detectable difference (MDD) was determined for each measured parameter and each gamma camera. The range of acceptable operation of a gamma camera system was set as the measured value +/- MDD for each QC parameter. The ability to detect and track small changes in the measured QC parameters was evaluated.[Abstract] [Full Text] [Related] [New Search]