Title: Effect of a polyethylene-lined administration set on the availability of diazepam injection. Author: Hancock BG, Black CD. Journal: Am J Hosp Pharm; 1985 Feb; 42(2):335-9. PubMed ID: 3976681. Abstract: Delivery of diazepam through a polyethylene-lined i.v. administration set and through a polyvinyl chloride (PVC) set was compared. Diazepam was prepared in concentrations of 50 mg/500 mL and 100 mg/500 mL in 0.9% sodium chloride injection and 5% dextrose injection in glass containers. Diazepam concentrations were measured by high-performance liquid chromatography at 0 through 5 hours in samples collected simultaneously from the glass solution containers and from the distal ends of a PVC administration set and a polyethylene-lined (non-PVC) set. Flow rates of 50 and 100 mL/hr were tested. For the non-PVC sets, diazepam concentration in the infusate was not significantly different from concentration in the glass container at any sampling time. The overall percentage of diazepam recovered was 100.7 +/- 6.8%. For the PVC sets, diazepam concentration in the infusate was less than in the container at all sampling times, and the overall percentage of diazepam recovered was 65.4 +/- 13.3% (significantly different from delivery for the non-PVC sets). Delivery through the non-PVC sets was not affected by flow rate, type of solution, or concentration of diazepam. For infusion periods of up to five hours, delivery of diazepam through polyethylene-lined i.v. administration sets was superior to delivery through polyvinyl chloride sets.[Abstract] [Full Text] [Related] [New Search]