These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Effect of a polyethylene-lined administration set on the availability of diazepam injection.
    Author: Hancock BG, Black CD.
    Journal: Am J Hosp Pharm; 1985 Feb; 42(2):335-9. PubMed ID: 3976681.
    Abstract:
    Delivery of diazepam through a polyethylene-lined i.v. administration set and through a polyvinyl chloride (PVC) set was compared. Diazepam was prepared in concentrations of 50 mg/500 mL and 100 mg/500 mL in 0.9% sodium chloride injection and 5% dextrose injection in glass containers. Diazepam concentrations were measured by high-performance liquid chromatography at 0 through 5 hours in samples collected simultaneously from the glass solution containers and from the distal ends of a PVC administration set and a polyethylene-lined (non-PVC) set. Flow rates of 50 and 100 mL/hr were tested. For the non-PVC sets, diazepam concentration in the infusate was not significantly different from concentration in the glass container at any sampling time. The overall percentage of diazepam recovered was 100.7 +/- 6.8%. For the PVC sets, diazepam concentration in the infusate was less than in the container at all sampling times, and the overall percentage of diazepam recovered was 65.4 +/- 13.3% (significantly different from delivery for the non-PVC sets). Delivery through the non-PVC sets was not affected by flow rate, type of solution, or concentration of diazepam. For infusion periods of up to five hours, delivery of diazepam through polyethylene-lined i.v. administration sets was superior to delivery through polyvinyl chloride sets.
    [Abstract] [Full Text] [Related] [New Search]