These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Quantitative measurement of 125I with NaI(Tl) scintillation survey meter]. Author: Nishizawa K. Journal: Radioisotopes; 1985 Feb; 34(2):78-82. PubMed ID: 3991923. Abstract: This paper describes a NaI(Tl) scintillation survey meter which can quantify 125I on a contaminated surface in situ. This survey meter has four inherent windows; 20-45 keV the singles photopeak region, 45-80 keV the sum photopeak region, 20-80 keV the total photopeak region, and 80 keV-. The activity was calculated based on the sum peak method. The calculated activity of a point source agreed with the standardized source activity within an uncertainty of about 10% both up to the distance of 3 cm from the detector surface along the axis of the detector and on the detector surface. The activity of a simulated plate source also agreed with the source activity within the same uncertainty.[Abstract] [Full Text] [Related] [New Search]