These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Comparison of a high performance liquid chromatographic and an enzyme immunoassay technique for quantitation of theophylline in serum.
    Author: Boeckx RL, Frith EM, Simons FE.
    Journal: Ther Drug Monit; 1979; 1(1):65-73. PubMed ID: 399811.
    Abstract:
    High performance liquid chromatograhic (HPLC) and enzyme immunoassay (EMIT) methods for quantitating theophylline concentrations in serum were evaluated. For both methods, standard curves were linear over the therapeutic range of serum concentrations. Precision was acceptable, with coefficients of variation being less than 9%. The HPLC assay was slightly, but not significantly, more precise. No interference was noted in either method by caffeine, theobromine, diphylline, 8-chlorotheophylline, or by metabolites such as 1-methylxanthine, and 3-methylxanthine, or by potentially interfering compounds such as urea and uric acid. The two methods correlated well, with a correlation coefficient of 0.98%. The enzyme immunoassay was superior in terms of lower costs, greater ease of performance, and potential for automation.
    [Abstract] [Full Text] [Related] [New Search]